What is Optical Photothermal IR (O-PTIR) spectroscopy
Co-propagating Mode
The O-PTIR technique overcomes the IR diffraction limit associated with traditional IR microscopy techniques by illuminating the sample with a mid-IR pulsed tunable quantum cascade laser (QCL) and measuring infrared absorption, indirectly with a visible laser beam.
When the QCL laser is tuned to a wavelength that excites molecular vibrations in the sample, absorption occurs, thereby creating photothermal effects, e.g., sample surface expansion and a change in refractive index. The visible probe laser, focused to sub-micron spot size, measures this photothermal response via a modulation induced in the scattered light, as shown in the video.
The IR laser can be swept through the entire range in under two seconds to obtain an IR spectrum.
Infrared resolution compared to the visible light probe
The illustration shows the variable spatial resolutions in common IR microscope systems over the traditional infrared range using two IR objectives of 0.5 or 0.7NA versus the constant spatial resolution of the O-PTIR technique using a 0.78NA objective (Kansiz, 2020). The O-PTIR technique provides wavenumber independent spatial resolution over the entire mid-IR range due to the use of a fixed wavelength probe beam at 532nm.
![Spatial resolution comparison chart-5](https://www.photothermal.com/wp-content/uploads/2024/04/Spatial-resolution-comparison-chart-5.png)
Spatial Resolution comparison between conventional FTIR, QCL-equipped IR microscope and the O-PTIR pump-probe infrared microscope
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
Counter propagating mode
An additional major innovation with the mIRage-LS platform is the addition of a new high spatial resolution IR mode, termed “counter-propagating” mode as seen in figure 3. The key enabler of higher spatial resolution is the ability to utilize regular high power glass objectives as the probe focus optic. Unlike “co-propagating” mode where all-reflective Cassegrain optics are used to focus both the IR pump and visible probe beam, the two beams are decoupled with the IR beam directed via the underside of the sample, allowing for high power glass objectives to be used to delivery and collect the probe beam.
The counter propagating method retains all the benefits of O-PTIR while also providing for increased spatial resolution and greater experimental flexibility, with options now for the use of water-dipping objectives, oil-immersion objectives, etc. Figure 4 illustrates some of these new modes of operation.
![Ultra high NA water-oil immersion objective with counterprogagating example-illustration copy](https://www.photothermal.com/wp-content/uploads/2022/12/Ultra-high-NA-water-oil-immersion-objective-with-counterprogagating-example-illustration-copy.png)
![Ultra high NA water-oil immersion objective with counterprogagating example-illustration copy](https://www.photothermal.com/wp-content/uploads/2022/12/Ultra-high-NA-water-oil-immersion-objective-with-counterprogagating-example-illustration-copy.png)
![Ultra high NA water-oil immersion objective - Horizontal - 2](https://www.photothermal.com/wp-content/uploads/2023/09/Ultra-high-NA-water-oil-immersion-objective-Horizontal-2-1024x292.png)
![Ultra high NA water-oil immersion objective - Horizontal - 2](https://www.photothermal.com/wp-content/uploads/2023/09/Ultra-high-NA-water-oil-immersion-objective-Horizontal-2-1024x292.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
What is Simultaneous IR and Raman spectroscopy
The visible probe laser of the mIRage also generates Raman scattered light from the same measurement area, which when collected, allows for the simultaneous acquisition of both IR and Raman spectra, with the same submicron spatial resolution, at the same location and at the same time. A schematic diagram for this optical beam path is shown below in figure 5. With this simultaneous measurement capability, researchers can utilize the complementary nature of both techniques and provide confirmatory spectral information for more accurate measurements.
Simultaneous IR and Raman spectral searching with 2D search result representation with KnowItAll®
With the advent of simultaneous submicron IR+Raman spectral acquisition, as seen in the schematic representation, not only are the IR and Raman spectra simultaneously collected, but now the spectral search of both IR and Raman spectral occur simultaneously, with a single click from the data acquisition software.
Furthermore, to aid in evaluating the results, the IR and Raman hit lists are presented in a 2D scatter plot, with the IR Hit Quality Index (HQI) plotted on one axis and the Raman HQI plotted on the other axis.
The figure below provides an example of a simultaneous IR+Raman search result for a common plastic. In such a 2D scatterplot, the best result is to be found in the upper right corner of the plot, where the result indicates both a high IR and high Raman HQI, thus providing for that confirmatory aspect as well as being complementary. The benefits of this unique approach are two-fold.
1. Productivity and ease-of-use gains, through the simple one-button spectral export and loading into KnowItAll® presenting the data to the analyst ready to be searched and answers delivered.
2. The increased accuracy and confidence in your data that comes with IR+Raman spectral providing both complementary and confirmatory data, i.e., IR search result can confirm the Raman and vice-versa.
![Mirage-OPTIR and Raman verticle new Raman Illustration](https://www.photothermal.com/wp-content/uploads/2022/12/Mirage-OPTIR-and-Raman-verticle-new-Raman-Illustration.png)
![Mirage-OPTIR and Raman verticle new Raman Illustration](https://www.photothermal.com/wp-content/uploads/2022/12/Mirage-OPTIR-and-Raman-verticle-new-Raman-Illustration.png)
![KnowItAll results image](https://www.photothermal.com/wp-content/uploads/2023/07/KnowItAll-results-image.png)
![KnowItAll results image](https://www.photothermal.com/wp-content/uploads/2023/07/KnowItAll-results-image.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
Co-located O-PTIR and fluorescence microscopy
mIRage-LS principal of operation: O-PTIR and fluorescence microscopy
The mIRage-LS provide an integrated Fluorescence camera in the optics path of the O-PTIR signal. This enables co-located O-PTIR and fluorescence microscopy.
Fluorescence (FL) images can be obtained by exciting the region of interest (ROI) with different visible wavelengths and detecting the radiation re-emitted at longer wavelengths. Fluorescence imaging has been used for a long time in both life and material science. The excitation and emission wavelengths of a given molecule are directly related to its inherent electronic states but would not affect O-PTIR data acquisition (Qin, 2020; Chen, 2022). As a result, widefield FL images can first map out the ROI before collecting the spatially resolved chemical information by submicron O-PTIR.
The system layout for a combined O-PTIR sub-micrometer IR microscope, co-located with fluorescence microscopy is illustrated in figure 7. In counter propagating mode, the pulsed IR beam is directed through the underside of the sample. The fluorescence and visible probe beam paths are co-located through the microscope objective.
Sub-micron IR spectroscopy with simultaneous Raman microscopy and co-located fluorescence microscopy
![How IR plus Raman plus Fluorescence Works-transparent image2a](https://www.photothermal.com/wp-content/uploads/2023/06/How-IR-plus-Raman-plus-Fluorescence-Works-transparent-image2a.png)
![How IR plus Raman plus Fluorescence Works-transparent image2a](https://www.photothermal.com/wp-content/uploads/2023/06/How-IR-plus-Raman-plus-Fluorescence-Works-transparent-image2a.png)
Microscope schematic for O-PTIR and co-located fluorescence microscopy
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
![Hairline faded ends 900px](https://www.photothermal.com/wp-content/uploads/2023/07/Hairline-faded-ends-900px.png)
FL-PTIR: How it works
FL-PTIR works by using temperature-dependent changes in fluorescent emission intensity to detect the subtle heating associated with chemically specific infrared absorption by a sample. Commercial fluorophore dyes, fluorescent proteins and naturally occurring fluorophores all have a strong temperature dependence in their emission efficiency (quantum yield), on the scale of ~1%/°C.
When a sample is exposed to pulses of IR light, IR absorbing regions of the sample heat up slightly and decrease the fluorescent emission from the heated regions. FL-PTIR uses a high-sensitivity s-CMOS camera to detect changes in fluorescent emission intensity from the sample in response to pulses of IR light from a wavelength tunable IR laser.
By comparing fluorescent emission camera images of the sample with and without the IR pulses, the mIRage-FL calculates widefield IR absorption images, while simultaneously capturing co-located fluorescence emission images.
Hyperspectral image arrays, i.e., an array of IR absorption images at different IR wavelengths, can be rapidly acquired in minutes. The hyperspectral arrays can then be analyzed to extract IR absorption spectra from any region of the image. IR chemical images and ratio images can be created by integrating the image stack over desired IR absorption bands.
![Fluorecsense diagram-new 4 text](https://www.photothermal.com/wp-content/uploads/2024/04/Fluorecsense-diagram-new-4-text.png)
![Fluorecsense diagram-new 4 text](https://www.photothermal.com/wp-content/uploads/2024/04/Fluorecsense-diagram-new-4-text.png)
FL-PTIR combines fluorescently labeled samples, with fluorescent excitation and pulsed IR light. The IR absorption by the sample reduces fluorescent emission intensity from sample due to fluorophore temperature sensitivity. The fluorescent emission intensity changes by 1%/°C which is 100X more than conventional OPTIR
![mIRage-logo-2017-900px](https://www.photothermal.com/wp-content/uploads/2018/10/mIRage-logo-2017-900px.png)
![mIRage-logo-2017-900px](https://www.photothermal.com/wp-content/uploads/2018/10/mIRage-logo-2017-900px.png)
![image of the mIRage: Combines FTIR and Raman infrared spectroscopy material analysis](https://www.photothermal.com/wp-content/uploads/2018/10/mIRage-product-1000px.png)
![image of the mIRage: Combines FTIR and Raman infrared spectroscopy material analysis](https://www.photothermal.com/wp-content/uploads/2018/10/mIRage-product-1000px.png)
Photothermal Spectroscopy Corp.
325 Chapala Street
Santa Barbara, CA 93101
Phone: (805) 845-6568
Email: info [at] photothermal.com